The FTUG-1600 is an ultrasonic gas flowmeter for natural gas and other compatible gases. Its multichannel design measures several flow-profile sections, helping reduce the influence of swirl and asymmetric flow while providing measurement redundancy.
A compact housing with integrated cable management supports robust installation and lower maintenance. The published size table covers DN25 to DN200. It is intended for urban gas, oil and gas, chemical, power and metallurgical gas-flow applications. Select the configuration and calibration range for the actual gas and operating conditions.
| Size | Mass (kg) | Flow range (m³/h) | Transition Qt (m³/h) |
|---|---|---|---|
| DN25 | 5.4 | 0.5–45 | 4.5 |
| DN32 | 6 | 1–70 | 7 |
| DN40 | 7 | 2–115 | 11.5 |
| DN50 | 7.7 | 3–180 | 18 |
| DN80 | 12.1 | 7–450 | 45 |
| DN100 | 18.6 | 12–750 | 75 |
| DN150 | 23 | 27–1700 | 170 |
| DN200 | 28 | 48–3000 | 300 |
| Nominal diameter | DN25–DN200 |
|---|---|
| Measurement | Multichannel ultrasonic gas flow |
| Transducer | Integrated stainless-steel probe, approximately 12 mm diameter |
| Maximum stated velocity | 70 m/s, depending on configuration |
| Maximum stated acoustic path | 2 m |
| Minimum stated probe temperature | −35 °C |
| Power supply | 3.6 V lithium battery, stated life ≥ 3 years; external 7–24 V DC supply requires a safety barrier |
| Insulation resistance | > 20 MΩ between external supply, signal output and enclosure |
| Media | Single-phase gases including natural gas, propane, LPG and air |
| Medium temperature | −25 to 60 °C |
| Nominal / operating pressure | ≤ 1.6 / ≤ 0.4 MPa |
| Ambient conditions | −25 to 60 °C; ≤ 90% RH |
| Pressure loss | < 1.0 kPa |
| Accuracy class | 1.0 |
| Indication error | Qt ≤ Q ≤ Qmax: ≤ ±1.0%; Qmin ≤ Q ≤ Qt: ≤ ±2.0% |
| Repeatability | Qt ≤ Q ≤ Qmax: ≤ 0.15%; Qmin ≤ Q ≤ Qt: ≤ 0.20% |
| Pulse output | Up to 5 kHz with external 7–24 V DC supply |
| Current output | Optional 4–20 mA, stated output error < 0.05 mA |
| Alarm output | Optional with external 7–24 V DC supply |
| Communication | RS485 / Modbus RTU with external 7–24 V DC supply |
| Display | Total volume, standard volume, operating volume, temperature, pressure and battery state |
| Connection / body | Flanged connection; aluminum-alloy body |
| Entity parameters — power terminals | Ui 28 V DC; Ii 93 mA; Pi 0.65 W; Ci 1.2 µF; Li 0.1 mH |
| Entity parameters — 4–20 mA terminals | Ui 28 V DC; Ii 93 mA; Pi 0.65 W; Ci 1.1 µF; Li 0.1 mH |
| Entity parameters — pulse/alarm terminals | Ui 28 V DC; Ii 93 mA; Pi 0.65 W; Ci 0 µF; Li 0 mH |
| Entity parameters — RS485/power terminals | Ui 7.5 V DC; Ii 100 mA; Pi 0.19 W; Ci 12 µF; Li 0 mH; verify the device approval and label before hazardous-area installation |